Two calibration Method Paint Thickness Gauge
Zoom Browse
Product Name: Two calibration Method Paint Thickness Gauge
Product Numbers: CTME-5302
Product Specifications: store 800 measurement data
Product Remark: Mil Gauge
Product Class:Measuring and Testing Instruments > NDT Instruments > Coating Thickness Gauges
Product Details
【DateTime:2018-11-10 Browse:】【Font: Big Mid Small】【Print】【Close
Two calibration Method Paint Thickness Gauge Paint Mil Gauge 0-1250μm
      Coating thickness gauge ,using test Magnetic thickness method, can be non destructively measuring magnetic metal matrix ( such as: steel, iron, nickel ) on non magnetic coatings thickness ( such as: galvanized, chrome, paint, electrophoresis, enamel, rubber, powder coating, enamel, etc.);using eddy current thickness method, can be non-destructively measuring non-magnetic metal substrate ( such as: aluminum, copper, stainless steel ) of non-conductive coating thickness ( such as: paint, powder, plastic, rubber, enamel, enamel, coating etc.)
Coating thickness characteristics
A.Fine small integrated, convenient to carry
B. Metal probe,Simple operation, quick test speed
C. Two test methods: a continuous test  and a single measuring method
D.Four statistical functions: maximum value, minimum value, average value, the number of measurements
E.Can store 800 measurement data
F.The instrument can be connected computers communicate to read the stored data to save or print (Application of U type host )
G.Two calibration methods: zero calibration and two point calibration
H.Operation process has the buzz sound prompt ( single measurement )
I.Automatic recognition of Fe and NFe substrates
J.Low voltage indication
K.Metallic test head, wear resistance, not easy to damage
L.Manual / automatic shutdown function
technical parameters
A.Measuring range:0-1250μm
B.Using environment: 0℃-50℃ and 20%RH—90%RH,not strong magnetic place.
C.the thinnest matrix: 0.4mm
D.measurement accuracy:±(1%-3%)+1.5μm
E.Resolution ratio:0.1μm /1μm
F.minimum curvature:凸bugle5mm;concave25mm
G.minimum matrix:10*10mm
H.power source:three alkaline batteries( no.7)
I.size:130mm*70mm*24mm
J.weight:100g
【Font: Big Mid Small】【Print】【Close】【Back to Top