Bench pH/mV/ISE/Conductivity/DO Meter
For multi-parameter high precision measurement laboratory measuring instrument, covering the pH, ion concentration, conductivity and dissolved oxygen, more than 10 kinds of electrochemical parameters. Suitable for petroleum, chemical, pharmaceutical, power plants and environmental protection industries in the field of education and research more widely used.
Meter meets with the requirement of international GLP standards. Its features up to automatic calibration, automatic temperature compensation, data storage, timing measurement, clock display, RS-232 output, function setting and self-diagnose etc.
There are three different measurement modes: stable display mode; time measurement mode; continuous measurement mode.
Automatically recognize 13 kinds of pH buffer solution. User can choose anyone from four series of standard buffer solutions: Europe & USA series, NIST series, CHINA series and Self-defined series
High purified water and ammonia added purified water pH measuring mode can be set up, especially suitable for electric power and petrochemical industry
High purified water and ammonia added purified water pH measuring mode can be set up, especially suitable for electric power and petrochemical industry
Automatically recognize 8 kinds of conductivity standard solution. User can choose anyone from two series of solutions: Europe & USA series and China series.
Adopt advanced and original conductivity measurement technology, use 1 pc of conductivity electrode (K=1), it can meet measuring accuracy requirement for range within 0.5μS/cm ~ 200mS/cm by using only one-point calibration
Automatically select conductivity measurement range, when measuring high pure water, equipment automatically enter into temperature compensation of nonlinear model, especially for electricity, microelectronics and the pharmaceutical industry requirements
Dustproof and waterproof meter meets IP54 standards. All sockets are seal protected by the silica gel caps.
Equips new type of DO electrode which has function of automatic temperature compensation and automatic salinity compensation as well as automatic air pressure compensation inside meter to make DO measuring mode has temperature, air pressure and salinity automatic compensation, easy to use, accurate to measure.
Polarogram type DO electrode has special electrode calibration cap. Only needs 3 to 5 minutes for electrode polarization. Electrode adopts combination diaphragm cap which very convenient to use. Each DO electrode equips with three spare diaphragms caps
There are 10 kinds of common ions can be selected in the ion concentration measuring mode, user also can self-define other ions.
Can switch 4 kinds of concentration unit : pX, mol/L, mg/L and ppm
pH
|
Measuring range
Resolution
Accuracy
Input current
Input impendance
Stability
Temp. compensation range
|
(-1.999 ~ 19.999) pH
0.1/0.01/0.001 pH
Meter:±0.002 pH, connect meter with electrode:±0.01pH
≤1×10-12 A
≥3×1012 Ω
+/-.002 pH/3h
(0 to 100) ℃( automatic or manual)
|
mV
|
Measuring range(mV/ORP/EH)
Resolution
Accuracy
|
-1999.9mV ~ 0 ~ 1999.9mV
0.1mV
+/-.0.03% FS
|
Ion Concentration
|
Measuring range
Display unit
Accuracy
Temp. compensation range
|
0.00 ~ 14.00 pX
pX、mol/L、mg/L、ppm
Meter:+/-.0.5 % reading(monovalent), +/-.1.0 % reading (bivalent)
0 to 60 ℃(automatic or manual)
|
Conductivity
|
Measuring range
Resolution
Accuracy
Temp. compensation range
Electrode constant
Reference temperature
|
Conductivity:(0.00~20.00) μS/cm (20.0~200.0) μS/cm
(200~2000) μS/cm (2.00~20.00) mS/cm
(20.0~200.0) mS/cm (200~2000) mS/cm(K=10)
Resistivity:(0 ~ 100) MΩ·cm
TDS: (0 ~ 100) g/L
Salinity: (0 ~ 100) ppt
0.01/0.1/1μS/cm 0.01/0.1/1 mS/cm
Meter:+/-.0.5% FS,connect meter with electrode:+/-0.80% FS
(0 ~ 50) ℃(automatic ) 0.1 / 0.5 / 1 / 5 / 10 / 50 / 100 cm-1
25℃、20℃ and 18℃
|
Dissolved Oxygen
|
Measuring range
Resolution
Accuracy
Response time
Residual current
Temp. compensation range
Salinity compensation range
Air pressure compensation
Automatic calibration
Electrode type
|
(0 ~ 40.00) mg/L(ppm) (0 ~ 200.0) %
0.1/0.01 mg/L(ppm) 1/0.1 %
Meter:+/-.0.10 mg/L, connect meter with electrode:+/-.0.40 mg/L
≤30 s(25℃, 90% response) ≤0.1 mg/L
(0 ~ 45) ℃(automatic) (0 ~ 45) ppt(automatic) (80 ~ 105) kPa(automatic) Air saturated by water; water saturated by air
Polarogram type
|
Temperature
|
Measuring range
Resolution
Accuracy
|
-10℃~ 110℃
0.1℃
5~ 60℃range:+/-.0.4℃ other range:+/-.0.8℃
|
Other Technical Parameters:
|
Data storage
Storage content
Power
Communication connector
Size and weight
Quality and safety certification
|
300 groups
Measuring value series number, measuring value, temperature, ATC or MTC state, measuring date and time
DC9V/300mA
RS232
160 × 190 × 70mm/880g
ISO9001:2008, CE and CMC
|
Working Condition
|
Environment temperature
Environmental humidity
IP rating
|
5 ~ 35 ℃(0.01grade) 15 ~ 30 ℃(0.001grade)
≤75%
IP54 Dustproof and waterproof
|
1. Model pH/mV/ISE/Cond/DO meter |
1set |
2. flexible electrode holder |
1pc |
3. intelligent stirrer |
1set |
4. plastic pH/ATC three-in-one combination electrode |
1pc |
5. glass pH combination electrode |
1pc |
6. temperature electrode |
1pc |
7. conductivity electrode |
1pc |
8. DO electrode |
1pc |
9. double salt bridge reference electrode |
1pc |
10. pH standard buffer solution(pH4.00、pH6.86、pH9.18/50ml) |
1btl/each |
11. 1413μS/cm conductivity standard solution (50 mL) |
2btls |
12. 9V power adapter (meter use) |
1pc |
13. 6V power adapter (stirrer use) |
1pc |
14 membrane cap (DO electrode use) |
3pc |
15. DO electrode inner solution (30 mL) |
1btl |
16. Cathode polishing paper |
2pcs |
17. stirring bead |
3pcs |
18. AA battery |
4units |
19. RS-232 communication cable |
1pc |
20. communication software disk |
1pc |
21. Operation manual |
1unit |
22. Brief operation instruction |
1unit |
23. Packing List |
1p |