PH ORP Temperature Meter Tester 3 in 1
Zoom Browse
Product Name: PH ORP Temperature Meter Tester 3 in 1
Product Numbers: PHME-ORP-3569
Product Specifications: ORP:-1999-1999mV
Product Unit: Temp:0-50℃
Product Remark: China ORP meter manufacturer
Product Class:Measuring and Testing Instruments > Environmental Test > ORP Meter
Product Details
【DateTime:2021-09-09 Browse:】【Font: Big Mid Small】【Print】【Close

PH ORP Temperature Meter Tester 3 in 1

Specification:

Measuring range:
PH:0.00-14.00 pH, ORP:-1999-1999mV, Temp:0-50℃
Resolution:PH:0.01pH, ORP:1mV, Temp:0.1℃
Accuracy:PH:±0.1pH, ORP:±5mV, Temp:±1℃
Automatic Temperature Compensation:0℃-50℃
Battery Type:4×1.5V(AG13)(Not Included)
Operating Temperature:0℃-50℃
Calibration: Manual,at one point through trimmer
Dimensions:188mm×35mm×35mm
Weight:98g

Operation:

1.Remove the protective cap.Turn on the meter.
2.First rinse the electrode with distilled water, and suck it with filter-paper.
3.Immerse the electrode in the solution to be tested.
4.Stir gently and wait for the reading to stabilize. Do not tap it against the glass.
5.Press the "pH/ORP" button, choose the PH or ORP for measuring.
6.Press the "C/F"button to switch between Celsius and Fahrenheit.
7.After use,turn off the meter. Rinse the electrode wth distilled water to minimize contamination.
8.Always replace the protective cap after use.

PH Calibration:

1.Rinse the ph electrode first with distilled water,and suck it with filter paper.
2.Turn on the meter. Press the "pH/ORP"button,choice the PH for measuring.
3.Immerse the ph electrode in pH6.86(or pH7.00)buffer solution,gently shake the meter and wait for a stable display.
4.Regulate the trimmer with the screwdriver provided until display shows "6.86"("7.00")at buffer solution temperature of 25℃.
5.Rinse the ph electrode with distilled water,and suck it with filter paper (not included).
6.Immerse the pH electrode in pH4.01 or pH9.18(pH10.0)buffer solution,gently shake the electrode and waiting for the reading to stabilize. The showing value and buffer solutions ph value is within error range.
7.The calibration of the instruments pH range is now complete.
Important:
The ph meter should be recalibrated under following situations;
The meter has been used for long time since the last calibration or use
The meter has been used in particularly continual conditions.
The utmost accuracy is required.

ORP Calibration:

1.Turn on the meter.Press the"pH/ORP" button,choice the ORP for measuring.
2.Rinse the electrode with distilled water.
3Immerse the electrode into the fresh 256mV ORP standard solution,then stir it gently,
4)Allow the reading to stabilize and with a small screwdriver turn the calibration trimmer until the display shows"256".
5)Rinse the electrode with distilled water and suck it with filter paper.
6)Switch off the meter.Replace the battery compartment cap.
The calibration of the instruments ORP range is now complete.

Electrode replacement:

1.Remove the protective cap.
2.Unscrew the plastic ring on the top of the electrode.
3.Pull out the electrode and replace it with a new one.
Make sure the gaskets are in place before screwing back the ring.