Portable pH/Conductivity Meter
Measurement of parameters such as pH, mV, conductivity, TDS, salinity, resistivity and temperature of water solution, the most cost effective portable analytical instrument
Main Features:
1. Automatic identification of 8 conductivity standard calibration solutions, two options of standard calibration solutions: Europe and America series and China series.
2. Meter meets IP57 rating, equipped with calibration solution and portable case.
3. Double-parameter meter, equipped with pH electrode and conductivity electrode, automatic identification of electrode by meter.
4. Able to set pH measurement modes of high purity water and pure water with ammonia.
5. Adopt advanced conductivity measurement technology, use conductivity electrode of constant K=1, only need one point calibration and meet the measurement precision requirement of 0.5μS/cm ~ 200 mS/cm.
6. Ability to switch among conductivity, TDS, salinity and resistivity. Multinomial calculation for TDS and salinity to ensure the conversion precision of the full scale.
7. Conductivity measurement mode has the function of non-linear temperature compensation for high purity water.
8. Built-in microprocessor chips, intelligent functions such as automatic calibration, automatic temperature compensation, data storage, function set-up, self-diagnose information, automatic power-off and low voltage display.
9. Automatic identification of 13 pH standard buffer solutions, three options of standard buffer solutions: Europe and America series, NIST series and China series.
Specifications:
Category |
Product Name |
Basic features and specifications |
PH |
Measuring range
|
(-2.00 ~ 19.99)pH |
Resolution
|
0.1/0.01 pH |
Accuracy
|
Meter:+/-0.01pH; Overall: +/-0.02pH |
Input current
|
≤2*10-12 A |
Input impendance
|
≥1*1012 Ω |
Stability
|
+/-0.01 pH/3h |
Temp. compensation range |
(0 ~ 100)℃(automatic) |
ORP |
Measuring range (mV/ORP/EH)
|
-1999mV ~ 0 ~ 1999mV |
Resolution
|
1mV |
Accuracy
|
Meter:+/-0.1% FS, Overall: +/-15mV |
Conductivity |
Measuring range |
Conductivity:(0.00~19.99)μS/cm (20.0~199.9)μS/cm (200~1999)μS/cm (2.00~19.99)mS/cm (20.0~199.9)mS/cm TDS: (0 ~ 100)g/L Salinity (0 ~ 100)ppt Resistivity:(0 ~ 100)MΩ·cm |
Resolution
|
0.01/0.1/1μS/cm 0.01/0.1 mS/cm |
Accuracy |
Meter:+/-1.0% FS, Overall:+/-1.5% FS |
Temp. compensation range
|
(0 ~ 50)℃(manual) |
Electrode constant
|
0.1 / 1 / 10 cm-1 |
Benchmark temperature
|
25℃,20℃ and 18℃ |
Other Technical Parameters |
Data storage
|
200 groups |
Storage content |
Series number, measuring value measuring unit and temperature, |
Power
|
Two AA batteries (1.5V *2) |
Size and weight
|
meter:(65×120×31)mm/180g: Suitcase: (360×270×76)mm/1.6Kg |
Quality and safety certification |
ISO9001:2000, CE and CMC |
Working Condition |
Environment temperature
|
5 ~ 35 ℃(0.01grade) |
Environmental humidity
|
≤85% |
IP rating
|
IP57 Dustproof and waterproof |
Standard package:
1. Model 723 pH/mV/Conductivity meter 1unit
2. 201T-S plastic pH/ATC three-in-one combined electrode 1pc
3. 2301-S plastic conductivity electrode 1pc
4. pH standard buffer solution(pH4.00, pH7.00 and pH10.01/50mL) 1btl/each
5.1413μS/cm conductivity standard solution (50mL) 1pc
6.Screw driver 1pc
7. Spare AA batteries 2pcs
8. Operation manual 1pc